The first block sets up the process for the entire life cycle. Top management provides resources and competent people, defines the policy and criteria for acceptability, and reviews the process. Each device has a risk management plan and a risk management file.
Risk analysis covers the intended use and reasonably foreseeable misuse, the characteristics related to safety, hazards, hazardous situations and the estimation of risk. Risk evaluation then compares the results against the acceptability criteria.
Risk control selects measures, implements them, verifies they work, considers any new risks they create and evaluates residual risk, including overall residual risk across the device.
A review before release confirms the plan was carried out. Production and post-production activities then collect information and feed it back into the file.